Operation installation parts list – Amico Gas Control Panel User Manual

Page 4

Advertising
background image

4

Amico Pipeline

1. Turn the Shut-off valve to the “Open” position.

2. Verify that the correct supply pressure is available on the supply gauge.

3. Attach a tool to the DISS outlet and tighten the nut (do not over tighten) or HP “Safety-Swing Coupling” outlet.

4. Adjust the pressure regulator to the required pressure by observing the outlet gauge.

5. Operate the surgical tool and observe the outlet pressure and adjust if needed.

6. When the control panel is not in use, reduce the outlet pressure to “zero” and turn off the supply valve.

Mount the back-box into the wall and adjust the adjustable side mounting brackets so that the top of the box will be

flush with the finished drywall. Connect the inlet pipe to the gas pipeline and the outlet to additional outlets (optional).

The pipeline can now be pressure tested to the required testing pressure. The outlet high pressure hose is shipped with

one end connected to the outlet and the other end connected to the inlet on the nitrogen back box. This will enable the

installer to pressure test the complete pipeline, including remote outlets, without having to attach the front assembly.

To attach the control panel front assembly to the control panel back box, you must first disconnect the braided hose

from the male PT adapter that is attached to the inlet pipe on the control panel back box. Then loosen and remove the

panel cap from the male PT adapter that is marked “output” on the back of the control panel front assembly. Attach the

braided hose from the back box to the male PT adapter (avoid over tightening). Connect the “input” braided hose from

the back of the control panel front assembly to the “input” of the male PT adapter in the back box (avoid over tightening).

Mount the control panel front assembly to the back box with the enclosed machine screws.

Nitrogen control panel English DISS (Relieving type)

N-CONP-E-REL

Nitrogen control panel English DISS (Non-Relieving type)

N-CONP-E

Instrument Air control panel English DISS (Non-Relieving type)

N-CONP-E-I

Instrument Air control panel English DISS (Relieving type)

N-CONP-E-REL-I

Operation

Installation

Parts List