QX100 Droplet Reader and QuantaSoft Software Instruction Manual

| 5

Chapter 1 QX100 Droplet Digital PCR System

QX100 Droplet Reader and QuantaSoft Software Instruction Manual

Using the QX100

Droplet Reader

2

1. Power on the QX100 droplet reader using the switch at the back. Allow

it to warm up for 30 min, then switch on the PC and launch QuantaSoft

software.

2. Check the indicator lights on the front of the droplet reader (Table 2.1). The

first two lights at left should be solid green, indicating power is on, there is
sufficient oil in the designated oil reservoir, and there is <700 ml in the waste
bottle. If the lights are flashing amber, the run cannot be started; clean out
the waste bottle or replace the oil (see Section 4.2.2).

Table 2.1. Status indicator lights on the QX100 droplet reader.

Solid green

Power on

Bottle levels OK*

Plate in place

Run complete

Flashing green

---

Oil <30% or waste >70%**

---

Run in progress

Flashing amber

---

Oil <10% or waste >90%***

---

Error during run

Off

Power off

---

No plate

Idle

* There is sufficient oil and room in the waste bottle to run 96 wells.
** The run can be started if <96 wells are run (for example, only 19% oil is required for 24 wells); if there is not enough oil

for the run, the software will not allow you to start the run.

***The software will not allow you to start the run.