7. Reconstitute the vial of standards in 10% serum-based standard diluent

(or a diluent similar to your sample matrix) by adding 781 µl of diluent.
Reconstitute the vial of control in 250 µl of the same diluent, as shown
below. Vortex at medium speed for 5 sec and incubate all vials on ice for
precisely 30 min.


8. Prepare a threefold standard dilution series and blank as shown below.

Vortex at medium speed for 5 sec between liquid transfers.

Note:

The controls are ready to use after reconstitution. No dilution is

needed. Controls are included with the fixed panel only.

9. After thawing samples, prepare according to the guidelines shown below.

10. Vortex coupled beads at medium speed for 30 sec and dilute to 1x in

Bio-Plex assay buffer as shown below. Protect from light.

# of Wells

20x Beads, µl

Assay Buffer, µl

Total Volume, µl

96

288

5,472

5,760

150 150 150 150 150 150 150 150

S1

S2 S3 S4 S5 S6 S7 S8

Blank

Diluent, µl

75

75 75 75 75 75 75

Transfer Volume, µl

Reconstituted

Standard

Sample Type

Diluent for Standards and Controls*

Add BSA

Serum and plasma

Sample diluent HB

None

Culture media, with serum

Culture media

None

Culture media, serum-free

Culture media

To 0.5% final

Lavage, lysate, other fluids

Sample diluent HB

To 0.5% final

* If using diluents other than the standard diluent HB provided, then users must establish

their own control ranges.

Sample Type

Dilution Factor

Diluent

Serum and plasma

MMP: 1:10 dilution

Sample diluent HB

Fluids

MMP: 1:4 and 1:40

Diluent + 0.5% BSA w/v

Bio-Plex Pro Human MMP Assays Quick Guide

Advertising