Bio-Rad Bio-Plex Pro™ Human Chemokine Assays User Manual

Bio-plex pro, Assays, Chemokine quick guide

Advertising
background image

This guide can be used to prepare and run a full 1 x 96-well assay plate.
For more information on a given step, refer to the complete instruction
manual. New users can download the manual, which includes detailed
instructions and a list of kit components, at www.bio-rad.com/bio-plex.

Initial Preparation

1. Plan the plate layout.

2. Start up/warm up the Bio-Plex

®

system (30 min).

n

Bring the 10x wash buffer, assay buffer, and diluents to room

temperature (RT). Keep other items on ice until needed

n

Begin to thaw frozen samples

3. Prime wash station for flat bottom plate or set vacuum manifold to

–1 to –3" Hg for filter plate.

4. Calibrate the Bio-Plex system by following the prompts within the

Bio-Plex Manager

™

software. This can be done now or during an

assay incubation step.

5. Prepare 1x wash buffer. Mix 10x stock by inversion to ensure all salts

are in solution. Then dilute 1 part 10x wash buffer with 9 parts dH

2

0.

IMPORTANT!

Pay close attention to vortexing, shaking, and incubation instructions.

Deviation from the protocol may result in low assay signal and assay variability.


For use with

Instruction Manual #

Bio-Plex Pro Human Chemokine Assays 10031990

Bio-Plex Pro

™

Assays

Chemokine Quick Guide