Bio-Rad Human Metabolic and Hormone Assays User Manual

Page 3

Advertising
background image

3. Add the appropriate amount of the standard diluent into the labeled

tubes according to the table below (this will be sufficient for duplicate
standard curves and blanks).

Standard

Volume of Standard Diluent, µl

Volume of Standard, µl

S1

––

150 from reconstituted vial

S2

100

50 of S1

S3

100

50 of S2

S4

100

50 of S3

S5

100

50 of S4

S6

100

50 of S5

S7

100

50 of S6

S8

100

50 of S7

Blank

100

––

4. Prepare working standards (S2–S8) by serial dilution. Transfer the

appropriate volume of standard into each of the labeled tubes with
standard diluent, as outlined above.

5. Vortex each standard at a medium setting before proceeding with the next

serial dilution. Change pipet tip at each dilution step.

6. The Blank tube consists of standard diluent alone.

C. Sample Preparation

1. Centrifuge serum or plasma samples at 1,000 x g for 15 min at 4°C to

remove particulates from all samples prior to use.

2. Prepare sample dilutions in 0.5 ml or 1.0 ml polypropylene tubes, as

required for the assay.

Bio-Plex Pro RBM Metabolic and Hormone Assays Quick Guide

Bio-Plex Pro RBM Metabolic and Hormone Assays Quick Guide

3. Dilution scenarios provided below are sufficient to run each sample in
duplicate.

Panel

Sample

Dilution

Volume of Sample, µl

Volume of Sample

Buffer, µl

Metabolic panel 1

1:5

20

80

Metabolic panel 2

1:5

20

80

Metabolic panel 3

1:500,000

(a) Prepare 1:50 5
(b) Prepare 1:100 5 of (a)
Prepare 1:100 5 of (b)

245
495
495

Metabolic panel 4

1:500

(c) Prepare 1:10 10
Prepare 1:50 10 of (c)

90

490

Hormone panel 1

1:5

20

80

Note: Controls are ready to use after reconstitution. No further dilution
is needed.

D. Dispensing of Reagents

1. Add 10 µl of blocker to all wells of the plate.
2. Add 30 µl of the standard, control, sample, or blank to the appropriate

well of the plate.

3. Vortex the capture beads at medium speed for 10–20 sec. Add 10 µl of

the beads to all wells of the plate.

4. Cover plate with plate seal and protect from light with aluminum foil.

Incubate on shaker at 850 ± 50 rpm for 1 hr at RT.

5. Wash the plate three times with 100 µl 1x assay buffer.
6. Vortex the reconstituted detection antibodies at medium speed for

10–20 sec. Add 40 µl to each well.

7. Cover and incubate at 850 ± 50 rpm, as in step 4, for 1 hr at RT. Do not

aspirate after incubation.

8. Prepare the required dilution of streptavidin-PE (SA-PE), as outlined in the

following table.

Note: Volumes in the table are for an entire 96-well plate. Smaller volumes
can be prepared, provided that the dilution ratios are maintained.

Advertising