UDM 35/40 PANEL METER USER MANUAL


8

Scaling the Inputs


When

is displayed, access is available for configuring the

electrical input range, the decimal point position and display span as
follows:

Electrical Input Range (HiE and LoE)

Allows the operator to define an electrical input range different from the
standard range setup at the “inP” programming step. As example, a UDM
with a full-scale 20mA (19.99mA) setting, it is possible to select an
electrical input range from 4.00mA (Lo.E) to 19.99mA (Hi.E) by
proceeding as follows:

When

is displayed, press

. “LoE” will appear.






Using the

key, toggle to select a digit, then press the

or

key to

choose the desired value. Using the

key, repeat as necessary until all

digits have proper values. Press

.


“HiE” will now appear.


Using the same procedure as above, toggle through the digits using the

key and choose the desired values with the

or

key. Press

to

confirm and advance to the next parameter, Decimal Point Position.