4. Invalid Command. If the command received from the host is not valid then the unit may return the following:-
<ACK>?A<CR>

Where:
<ACK> is Acknowledge (6 Dec, 06 Hex)
? is the character “?” (63 Dec, 3F Hex)
A is the responding units address
<CR> is a Carriage Return (13 Dec, 0D Hex)
If the address received from the host does not match the units address then the unit will not respond at all.

7 SPECIFICATIONS

7.1 Technical specifications

Input:

Ratiometric 4 wire strain gauge.

Input Sensitivity:

80Ω to more than 2000Ω

Excitation:

5VDC

Accuracy:

Up to 0.005% of full scale for alarms and display, depending on sample rate etc., see
resolution table which follows. Using and

calibration method accuracy is 1%

Sample rate:

5 to 100 sample per second, selectable.

ADC Resolution:

Up to 22 bits depending on sample rate and mV/V input, see 7.4

Display update:

Up to 4 per second, apparent update varies with

setting

Conversion Method:

Sigma delta

Microprocessor:

HC68HC11F CMOS

Ambient temperature:

0 to 50o C

Humidity:

5 to 95% non condensing

Display:

4.5 digit LCD 11.7mm digit height display range -19999 to 19999

Power supply:

Battery 9VDC type 216 (alkaline recommended)

Battery life:

Typically 15 hours when switched on with 350 load cell connected

Low battery warning:

message plus beeper

Output (standard):

RS232 serial communications

7.2 Options

External memory:

Memory chip, 9 pin D connector and shell

7.3 Physical characteristics

Dimensions:

80mm (w) x 145mm (l) x 39mm (d)

Connections:

9 pin “D” connector (female at instrument end)

Weight:

250gms including battery

7.4 Resolution table

Effective resolution (bits) for DFI 200 over full scale – mV / V input