1. No wire calibrations are stored of the same diameter as the wire you are looking to measure, and
2. You do not have sheaves of the same diameter.

If both of these conditions exist, contact Cooper Instruments.

Contact Cooper Instruments to improve accuracy for a specific wire type by calibrating to it.

5.3 Loading Error

A tensiometer works by deflecting the cable, which makes the cable path longer than when a tensiometer
is not installed. When the tensiometer is removed, the wire tension decreases as the cable length is
restored. This effect is known as loading error. The WTM 502 design elongates the cable by a mere 0.08
inch (2mm), making loading errors extremely small.

5.4 Non-repeatability

The WTM 502’s sheave with bearing design provides the best mechanical performance. It is also superior
at detecting tension that is being added or removed.

5.5 Non-linearity

Most three-point tension meters employ only linear characterization and have large errors at the
midpoints (up to 15%). The WTM 502 uses multi-point segmenting to correct for non-linearity, reducing it
to less than 0.2%.

5.6 Wire Characteristics

Creep

Every material including steel exhibits creep under load. It will neck down over time,
quite quickly over the first few seconds and much slower as time progresses. A wire
cable also sees creep from the wire spacing and wind. This effect is seen as a
display that drifts lower after it has been clamped in line.


Variations

Material that varies in diameter or shape will have different output at the same
tension


Strands

The best cable assembly is one that is perfectly round, as it will not change contact
geometry with the wire twist. The closer the wire cable cross section appears to be
round, the better the measurement performance will be.


6.0 TROUBLESHOOTING

Problem Possible

Cause Solution

Powers on momentarily and
turns off

Low battery

Replace with high quality alkaline
batteries

Low battery

Replace with high quality alkaline
batteries

Batteries installed backwards or no
spring contact

Insure that positive terminal of
both batteries (nub) face inward-
towards the black cap. Check
that spring is attached to the
battery cap.

Does not power on

Software reset

Remove battery cap and reinstall
after one minute. Attempt to turn
poser on again

CF 142