Kangaroo, Enteral feeding pump – COVIDIEN Kangaroo™ 924 Hang Tag User Manual
Page 2
 
Kangaroo
™
924
Enteral Feeding Pump
Troubleshooting Guide
15 Hampshire Street
Mansfield, MA
02048
1-800-962-9888
508-261-8000
www.covidien.com
SYMPTOM PROBABLE CAUSE
CORRECTIONS
Flow Error 
Alarm
(FLo Err)
Feeding container is empty.
Feeding tube or feeding set 
tubing is occluded. 
 
Drip chamber is improperly 
placed in pump.
Drip chamber walls are coated 
with feeding formula. 
Sensors in upper drip chamber 
guide are blocked or dirty. 
Drip chamber is too full.
Refill feeding container. If bag has been in use for 24 hours, 
replace.
Locate point of occlusion and correct. Note: occlusion may 
occur upstream or downstream from pump. Check to make 
sure roller clamp on feeding set tubing is open. Check the 
patency of feeding tube.
Check to make sure drip chamber is properly “locked” in 
position in drip chamber guide.
Check to make sure formula is not preventing sensors from 
proper operation. If formula cannot be removed from inside 
walls of chamber via chamber manipulation, replace set.
Check to make sure sensors are free of dried formula. 
Remove any deposits by using a cotton swab dampened 
with warm soapy water.
Low 
Battery 
Alarm 
(Lo bAt)
Battery has been run down 
below point of maintaining 
accuracy of pump.
Plug pump power cord into wall outlet.
Hold Error 
Alarm 
(HLd Err)
Pump has been left in the hold 
mode for over 5 minutes.
Press to silence alarm, then press again to stop 
flashing, then press again to restart pump.
No Set 
(LoAd SEt)
Pump set has been improperly 
placed onto pump.
 
 
 
If "dots" appear with message; 
too much ambient light
Check to make sure only a Kangaroo 924 Pump Set has 
been placed onto the pump.
Check to make sure that the optical safety sensor on pump 
set has been properly positioned into the safety sensor lock on 
pump. Make sure there is no water or other contaminants 
on the safety sensor or safety sensor lock window. 
Move Pump away from bright light.
System 
Error Alarm 
(SYS Err)
Rotor is not turning while 
pump is running.
Rotor is turning but is not 
detected by rotor sensors.
Return pump for technical service.
Return pump for technical service.
No Audible 
Alarm
Battery is fully discharged.
Alarm circuitry failure.
Charge battery for a minimum of 12 hours.
Return pump for technical service.
START
HOLD
START
HOLD
START
HOLD