COVIDIEN Genius™ 2 Tympanic Thermometer Checker-Calibrator User Manual

Page 7

Advertising
background image

Genius

TM

2

Checker/Calibrator

Table of Contents

5

Section VI - Operational Procedure

The following procedure will describe the steps required to operate the Genius 2 Checker/Calibrator. In order to facilitate familiarity

with the product, it would be advisable to use this section of the manual during the calibration check of the first few thermometers.

Installation of USB Flash Drive (Optional)

Install a USB flash drive into the USB port located on the front panel if a report file written to the USB flash drive is desired. If the USB

flash drive is installed after performing a calibration check, an error message may occur when attempting to write the report file if

the operator does not wait for the USB flash drive to complete initialization. If the Genius 2 Checker/Calibrator is powered off, all

internal memory (not yet saved to the USB flash drive) will be lost.

Apply Power

Make sure the AC-adapter end of the power module is plugged into a grounded outlet. Connect the power module to the front panel

power connector. Apply power to the Genius 2 Checker/Calibrator by using the power switch on the front panel.

Start-up Screen

The first screen on the LCD display is the Initialization Screen, which displays the software revision and date of the Genius 2

Checker/Calibrator. If there is an internal problem during initialization, an Error screen will be displayed (See Section VII).

Reminder Screen

The next screen on the LCD display is a reminder to let the Genius 2 Checker/Calibrator warm-up for at least 15 minutes after

applying power. This will let the infrared blackbody targets warm up to the proper temperatures. To exit this screen, press the

encoder knob.

Main Menu

The Main Menu screen is displayed next. This screen has four menu options:

Check Calibration

Preferences

Set User/Site Information

Set Time/Date

To enter any menu option, use the encoder knob to scroll over and highlight the desired menu option, then press the encoder knob.

Preferences

The Preferences menu is used to setup the language option. Press the encoder knob to enter the Language Selection screen.

Language Selection

The Language Selection screen lists the available languages for the LCD screen text. To select a language, scroll over the selection and

it will be highlighted. Press the encoder knob to complete the selection and return to the Main Menu screen. The selected language

will be stored in non-volatile memory and will be used next time the unit is powered on.

Set User/Site Information

The Set User/Site Information screen is used to enter the operator’s name and the organization and/or site information.

This information will be written to the test report file.

To enter the information, use the encoder knob to scroll through the alphabet for a given letter position in the operator name or site

information. When the correct letter is shown, press the encoder knob to select the letter and advance to the next letter position.

To enter a space, just press the encoder knob when a space or blank letter is shown in that letter position. The user name can be up

to 20 letters long. Use the encoder knob to scroll to the organization field and enter a name for the site in a similar manner. To exit

the menu, scroll to the DONE text and press the encoder knob.