Gas measurement instruments ltd – Gas Measurement Instruments GT Series User Manual

Page 2

Advertising
background image

Switch Instrument ON: Press and hold RH button

, for one second, to

The battery capacity level is displayed
for approximately five seconds during
the warm-up cycle, then on the top of
the display during normal operation.

The time and date from the instrument’s
built-in clock is then displayed on the
screen during warm-up.

Next, the Calibration Due Date is displayed (if configured).

This feature has five options:

1.

Cal Due Date message is not displayed.

2.

Cal Due Date and overdue date messages are displayed.

3.

(Default) Cal Due Date message is displayed with user acknowledge if
overdue. If the Calibration date has expired, the user must acknowledge
that Calibration is overdue by pressing and holding the RH Button, as
indicated on the bottom line of the display. Alternatively, press and hold
the LH button to initiate the instrument shut-down sequence.

4.

Cal Due Date message is displayed with user acknowledge for extended
period option, if overdue. If the Calibration date has expired, the user will
have to either press and hold the RH Button to accept the extended
period, or press and hold the LH button to proceed with the instrument
shut-down sequence.

5.

Cal Due Date message is displayed with user shut-down if overdue.

Service Due Date is then displayed (if configured). This feature also has five
options similar to Cal Due Date and operate in the same way except that the
default is same as Cal Due option 1, i.e. Service Due Date is not displayed.

The instrument will now automatically select Leak Test Mode, as default.

For all other GT series instrument
features and button functions, refer to
User Handbook: GMI Part No. 67112

GAS MEASUREMENT INSTRUMENTS LTD.

Note: Not applicable to Confined Space Mode or Pressure Mode.

Configurable options are available
to either start up in any other
operational mode or, to start up in
the mode last used.

Check that instrument displays
sample / flow fault when inlet is
blocked (with a finger for example).

Switch Instrument OFF or Re-enter Mode Menu: Press and hold both

LH

and RH

buttons simultaneously to initiate shutdown sequence.

For the first two seconds, the Mode menu (shown above) will be displayed.
Keep buttons pressed for a further three seconds to complete OFF sequence.
Abort switching OFF at any time by releasing the buttons.
Note 1: In Carbon Monoxide and Barhole modes, a sub-menu may be
displayed before reaching mode menu.
Note 2: In Confined Space Mode, to prevent inadvertently switching
instrument OFF or changing mode while alarms are active, the user must
press and hold both LH and RH buttons simultaneously for an additional
five (5) seconds before the mode menu appears.

Change Operating Mode: (Only after purging in fresh air) Press UP

or DOWN

buttons to highlight and select required operating mode

then press and hold RH

button to confirm choice.

Invert Display: Press LH button

to invert the display and allow

ease of reading in awkward situations.

If fault is not displayed, check tightness of all fittings.

Pump: Press RH button

to switch the pump ON and OFF.

Range: Press Centre button

to change range. See table - over . .

switch the instrument ON in fresh air.
The instrument now begins its ‘30 second’ warm-up cycle, during which, a
countdown timer appears in the top right hand corner of the display. The display
backlight illuminates and remains ON until the warm-up cycle is complete and
then automatically switches OFF.

During the warm-up cycle, the instrument display identifies the model, serial
number, software version and battery status information as shown below.
(Note: By default, all GT series instruments are configured for Datalogging)

67130_04.pmd

1/23/2008, 9:39 AM

2