Gas Measurement Instruments PS200 Series User Manual

Ps200, Series quick operating instructions

Advertising
background image

PS200

Series

QUICK OPERATING INSTRUCTIONS

Switch ON

Press and hold the Right Hand button

for one second to switch the instrument ON and
initiate a warm-up routine.

The display will show the serial number,

software version, battery status and calibration
gas. Next, the date and time are shown, followed
by the Calibration Due date. If expired, press the

button to acknowledge.

Lastly, a sensor check is run. If sensors are

working OK and were zeroed correctly, a

symbol is displayed adjacent to the gas type. If a
sensor fails the zero check, the display will be
red, the audible / visual alarms activate and a
symbol is displayed adjacent to the faulty gas type.
To acknowledge this fault, press the

button.

Switch ON - Manual Bump Test

Press and hold the

button to switch the

instrument ON and initiate a Manual Bump Test
(after the warm-up routine is complete). When
concluded, the instrument will operate as normal.

Refer to Essential Health and Safety Requirement (EHSR) and PS200

Series User Handbook before first use.

Backlight / MAX & MIN

Press the

button once to switch the backlight

ON. (After 20 seconds it will automatically switch
OFF.) Press again to view MAX readings and a further
press to view the MIN readings. To reset MAX / MIN
readings, press and hold the

button for 2

seconds when either MAX / MIN screen is displayed.

Alarms Reset / Acknowledge

In alarm condition, the display will be red and the
audible and visual alarms are activated. When the
gas readings (of a Latching Alarm) are safe, press
and hold the

button to reset / acknowledge the

alarm. (If alarms are Non-Latching, the audible and
visual alarms reset automatically when gas readings
are safe.)

Normal Operation

The standard operating screen for the 4 Gas model
is shown here. Other models will report the relevant
gases they are capable of detecting.

Switch Pump ON / OFF

Press and hold the

button to toggle the pump ON / OFF

(Pumped instruments only).

Manual Datalog

Press the

button once to perform a manual datalog, briefly indicated on

the display as ‘LOGGING’.

Switch OFF

Press and hold both

and

buttons, for 3 seconds, to switch the

instrument OFF.

Advertising