Iris Sample Processing StatSpin® Express 3 Primary Tube Centrifuge User Manual

Page 13

Advertising
background image

9

Troubleshooting

No power light

↓

Is the line cord completely plugged

into the wall?

→ No→

Plug line cord into the wall.

↓

Yes

↓

Is the wall outlet working properly?

→ No→

Try another wall outlet.

↓

Yes

↓

Is the fuse blown?

→ Yes→

Replace fuse.

↓

No

↓

Contact Iris Sample Processing or

an Authorized Repair Facility.

Centrifuge will not spin, or shuts

off prematurely

↓

Is the cover closed and latched?

→ No →

Completely close the cover.

↓

Yes

↓

Is the error light flashing?

→ Yes →

Identify source of error from flashing
indicator lights, correct the error condition,
and push the Stop/Open button to clear the
error.

↓

No

↓

Contact Iris Sample Processing or

an Authorized Repair Facility.

Centrifuge will not open at end

of cycle

↓

Is the bowl gasket completely

seated and sealing the bowl?

→ NO→

Press the bowl gasket all the way down into
the housing slot.

↓

YES

↓

Contact Iris Sample Processing or

an Authorized Repair Facility.