1 maintenance at regular intervals – Baumer BA DenexEdgeDetectorII User Manual

Page 14

Advertising
background image

User Manual DENEX Edge Detector II

14/20

Baumer Electric AG

Version 2013-02, V1.6

www.baumer.com

Frauenfeld, Switzerland


Some products are not counted (too many copies in the bundles)

•

Check the mechanical installation. The sensor must be parallel to the stream.

•

Check the Application setting; DIP 5-8.

•

Check the interface. See previous page.

•

If the product stream is very irregular, it could be that the sensor is blocking out
products coming too close. Try choosing an application with a shorter blocking.
See chapter 4, ”SOFTWARE FUNCTIONS”.

•

Make sure that the stacker has not got an internal Blocking Zone which is set in a
way that it blocks out pulses from the sensor.

•

If the product with the problem is very black and very thin, try lowering the sensor 10-
20 mm closer to the product stream.

•

Are the products very tilted when they pass the sensor? The max angle allowed is
15

°

for a 0.25 mm product.











9.1 Maintenance at Regular Intervals

To ensure the best operation, by following the steps above:

•

Clean the glass aperture with alcohol to remove ink dust.

•

Remove any loose pieces of paper that can produce false counts

•

Check that the target plate is still mounted the way it should be