Baumer BA DenexGripperLED User Manual

Page 12

Advertising
background image

User Manual DENEX Gripper LED

12/24

Baumer Electric AG

Version 2013-02, V1.4

www.baumer.com

Frauenfeld, Switzerland

To adjust the sensor to get a blocking of approximately 50 % do the following:

•

Make a bracket that enables the sensor to be moved sideways, or order one from

Baumer. Specify type of conveyor!

•

Set the sensor in Gripper Pulse Mode, see chapter 6.4

•

Watch the Output LED during production. It should be flashing.

The next step can be dangerous! Watch your hands if you do the movement during
run!

•

Move the sensor sideways until the Output LED is ON for approximately as long as it is

OFF to get 50 % duty cycle.

•

You can also check with a scope, either on the Output directly or between TP8 and

TP1, see chapter 6.

If this is done OK, the sensor will only count one edge for half a Gripper, which will be the
ideal situation.