5

LABS

LABS Cryostorage Specifications

Liquid nitrogen at atmospheric pressure weighs 1.78 lb./liter (0.8 kg/liter). All Taylor-Wharton Gas Equipment
CryoScience systems are designed to support the full weight of liquid nitrogen and a complete stainless steel or
aluminum inventory control system with boxes and specimens.

Dimensions

20K 38K 40K 80K 94K

External Operating Height

in.

42

44

42

42

40

(Top of Step to Lid opening)

mm

1067

1118

1067

1067

1016

Step Height

1

in.

11

11.5 11

11/17 11/22

mm

279

292

279

279/432 279/559

Height

in. 53 56 53 53 63

(Floor to Opening)

mm

1346

1422

1346

1346

1600

Usable Height, Internal

in.

30

30

30

30

35.75

mm 762 762 762 762 908

Outside Diameter

in.

34

42

45

59.5

59.5

mm 864 1067 1143 1511 1511

Internal Working Diameter

in.

29.5

37.8

40.5

55

55

mm 750 959 1029 1397 1397

Neck Opening

in.

13

18

18

24.5

24.5

mm 330 457 457 622 622

Capacity

Total LN

2

Capacity

L

407

626

606

1350

1630

LN

2

Under Capacity

L

46

72

76

150

300

Power Supply

2

VAC

110/220 110/220 110/220 110/220 110/220

Weight, Empty

lb.

650 930 920 1550 1700

kg 295 422 417 703 771

Maximum Gross Weight

3

lb.

1375 2050 2000 3956 4600

kg

624 930 907 1794 2086

Inventory Control System

Specifications
No. 100/81 Cell Racks

14

26

30

58

60

No.

Shelves/Rack

13 13 13 13 15

No. 25 Cell Racks

4 12 8 12 8

No.

Shelves/Rack

13 13 13 13 15

Vial Capacity, 2 ml

4

19500 37700 41600 79300 93000

Bag 25 ml (7 Level Rack)

5

1624 2688 3108 5999 6856

Bag 50 ml (7 Level Rack)

5

924 1582 1876 3381 3864

Bag 250 ml (5 Level Rack)

6

500 900 990 1920 2304

Bag 500 ml (5 Level Rack)

6

420 690 830 1610 1932

Footnotes:

1

Model LABS 94K has an independent two stage stainless step. Also available as an option on the LABS 80K.

2

Controller reduces power to 24VAC or less. Power outlet configuration should be indicated at time of order.

3

Includes the empty weight and total LN

2

capacity weight.

4

2 ml vial size; 12.5 mm O.D. internal thread. 6-2 half-rack capacity of 94,200.

5

Model LABS 94K, 8 Level Rack

6

Model LABS 94K, 6 Level Rack

TW-LABS-Manual-Digital-TOC-BlackandWhite.indd 5