2. Accuracy

The overall probe accuracy is a combination of the thermistor's interchangeability
specification and the accuracy of the bridge resistor. The Steinhart and Hart
equation used to calculate temperature has a negligible error (Figure 2-1). In a
"worst case" the errors add to an accuracy of ±0.6

°C over the range of -40° to 70°C

and

±0.49°C over the range of -20°C to 70°C. The major error component is the

interchangeability specification (tolerance) of the thermistor. The bridge resistor
has a 0.1% tolerance with a 10 ppm temperature coefficient. Figure 2-2 shows the
possible worst case probe and measurement errors.