Theory of Operation

In SI Units, this becomes the formula where

R

NI

H

715

.

0

H is the magnetic field in amperes per meter

N is the number of turns per coil

I is the coil current in amperes

R is the coil radius in meters

If instead of magnetic field intensity (H) the magnetic flux density (B) is the

parameter of interest the equation, when expressed in CSG units,

becomes

R

NI

H

B

μ

μ

899

.

0

=

B is the magnetic flux density in gauss

µ is the relative permeability

N is the number of turns per coil

I is the coil current in amperes

R is the coil radius in centimeters