Earth line choke switch, Ac receptacle, Artificial hand – ETS-Lindgren 3810-2 LISN User Manual
Page 16

E
ARTH
L
INE
C
HOKE
S
WITCH
The safety ground isolation choke selector switch switches the 1.6 mH earth line
choke IN and OUT of the safety ground circuit. The ground choke is designed
and manufactured with sufficient capacity to conduct the maximum current rating
of the Model 3810/2 and at no time is the safety ground of the unit compromised.
The earth line choke avoids a double RF ground connection (safety ground and
measurement ground) in the conducted emissions test setup.
AC
R
ECEPTACLE
Connect the Equipment Under Test (EUT) to the panel-mounted AC receptacle.
The style of receptacle is determined by the model specified. Following are the
standard output receptacle types:
•
NEMA (Type 5-15R)
•
Schuko (Type CEE 7/7)
•
British Standard (BS 1363)
•
Australian (AS 3112)
A
RTIFICIAL
H
AND
In conformance with EN55014 and BS800, the artificial hand connection is used
to test handheld equipment that is provided without earth connections.
16 |
Installation and Application