Installation
|
31
The MAPS is designed to operate using 208–230 VAC single-phase 50 or 60 Hz
power.
The branch circuit supplying power to the motor bases must be protected from
excess current according to local electrical codes. Integral circuit protection is
provided in the motor base assembly.
Check that the conductor size is adequate for the motor load and the distance
from the mains source. Improperly-sized conductors will lead to a high voltage
drop in the power conductors and cause reduced starting torque and premature
motor failure.
The motor base assembly is provided with an IEC-320 power inlet for connecting
to the mains.
Prior to servicing the turntable or the turntable motor base,
remove the power connection for safety.
1. Connect the fiber optic control cable and install the power connection per local
electrical code. See the controller manual for information on connecting fiber optic
cables.
2. After the fiber optic cable is installed, secure it with a wire tie to one of the leveling
screws.