ETS-Lindgren HI-1501 Microwave Oven Survey Meter User Manual
Page 14

Page — 6
HI
n
1501 Manual
Parameter
Error (dB)
1 mW/cm
2
Calibration - Precision
- Accuracy
+0.42, -0.16
±0.09
Nonlinearity and AM response
±0.09
Near Field vs. Far Field
±0.29
Receiving Pattern
-0.11
Temperature Response
±0.17
Frequency Response *
±0.04
Polarization *
±0.21
RFI *
±0.04
Drift *
±0.02
Total
+1.27 / -1.12
* Errors combined in RMS manner
Calibration Method
Each meter is calibrated by placing the probe in a CW
(continuous wave) 2450 MHZ electromagnetic field
generated by a crystal controlled solid state source. The
source feeds an anechoic chamber through a section of
waveguide terminating in a slot radiator. Calibration is
performed at a level of 1 mW/cm
2
and a standard
temperature of 75
/
F.
The calibration is also checked at 5 mW/cm
2
. The
accuracy of the field is determined by comparing with an
LCR (local calibration reference) which is traceable to
NIST (National Institute of Standards and Technology