Operation
Before placing into operation, follow the safety
information in the ETS-Lindgren
Product Information Bulletin included with your
shipment.
The HI-2602 monitor and probe are calibrated as an integral unit. Care must be
taken not to interchange probes and monitors without recalibration. Operating the
HI-2602 with different probes may cause erroneous readings.
Front Panel Controls
RF Control—contains a switch for disabling the interlock and a red LED to
indicate when RF is over the preset RF fail level. A push button switch is used to
reset the system when the LED illuminates. A momentary toggle switch in
combination with a potentiometer is used to set the RF fail level.
Power—contains the main power switch and a green LED indicator for indicating
power on.
System—contains a potentiometer for adjusting the needle zero, a red LED
indicating power failure and a push button switch for testing the desired RF fail
level.
Analog Meter—indicates RF readings from 0 - 2 mW/cm².