39
9.0 Specifications
Frequency:
2450 + 0.1 MHz over an ambient
temperature range of 20 degrees to
24 degrees C
Frequency Stability:
50 ppm/°C + 0.0025%
Dynamic Range:
0.1
– 5 + mW/cm
2
Plane-wave equivalent power
density at probe test location.
Input Power:
100
– 240 VAC
0.8
– 0.4 A max
50
– 60 Hz
Fuses:
250 Volt, 1.0 Amp, Type T
(5 mm x 20 mm)
Output:
2450 MHz, 4 W max
Do not operate without 50 ohm load.
Local Calibration Reference:
HI-1710 Microwave Survey System
(HI-1710 LCR)
Size:
Width: 770 mm wide
Depth: 1040 mm deep
Height: 1230 mm high
(1360 mm with HI-2790B
Power/Control Module)
Weight:
45 kg (shipping)
Support:
Castor wheels for ease of location