ETS-Lindgren HI-3604 ELF Survey Meter User Manual
Page 35

HI-3604 Manual
Page -- 27
longer part of the waveform trace in Figure 5-6 is related to the
time it takes for the beam to travel from the top of the screen
to the bottom; the very short transition in the waveform is the
time it takes for the beam to return to the top of the screen
after reaching the bottom.
The peak-to-peak output (in millivolts) from the waveform jack
is nominally related to the observed field as follows:
Electric Field:
Range (full scale)
Max. Output
(nominal)
19.99 volts
20 millivolts
199.9 volts
200 millivolts
1.999 kilovolts
2 volts
19.99 kilovolts
200 millivolts
199.9 kilovolts
2 volts
Magnetic Field:
milligauss Range
10 mV/milligauss
Gauss Range
100 mV/Gauss
Refer to Section 7, Using the Fiber Optic Remote Control for
information regarding the "recorder" output.