GAUGE SPECIFICATIONS:
Pressure display range:

29 InHg to 600 psig (-98kPa to 4134 kPa)

(+/- 1 PSI to 200 psi, 0.5% to 600 psi)

Temperature display range:

-40 °F to 200°F (-40 °C to 93 °C)

Operating temperature:

-10°F to 120°F (-12 °C to 49 °C)

(+/- 1 °F * 32°F/120 °F, +/- 1.5 °F * -10 °F/32

°F)

Maximum overpressure:

800 psig (5512 kPa)

Battery life and type:

6 Months with two 9v Alkaline battery

Auto-shutoff time:

60 minutes (extended 60 minutes by any button touch)

24 hour mode also available
Refrigerant data source:

NIST REF PROP software and manufacturers data

Uses one or two 9v Alkaline batteries

(Two batteries included)

Batteries can be replaced by taking

off the back of the box

LOW SIDE DISPLAYS

PRESSURE, SUPERHEAT,

TEMPERATURE (T1) and SATURATION
TEMPERATURE

PUSH to toggle between displays

PUSH once to display MAXIMUM,

MINIMUM and AVERAGE values of
current displays

HOLD DOWN to clear memories

Use Temperature Jack (T1) for

Superheat Testing

APPLICATION TIPS:
Zero Before Pressure Is Applied

Allow the manifold to Zero at Turn-On: The manifold displays
will zero each time the manifold is turned on without
pressure. Zeroing the gauge compensates the Pressure
display for changes in (1) Altitude and (2) Barometric pres-
sure.

Pressure Calibration

Don't be alarmed if your digital manifold does not agree with
your mechanical gauges. The digital manifold is calibrated
with a very accurate pressure and is not affected by vibration,
motion or position.

Extending On-Time

The digital manifold will turn off automatically after 60
minutes to save battery life. If any button is touched, the

manifold will stay on for another 60 minutes.
The on time can be extended to 24 hours by pushing the ON-
OFF button 2 times quickly.

Over/Under-Range Indicator

Pressures or temperatures below or above the rated

ranges will cause a “1” to be displayed.

Low Battery Indicator

Low batteries will be indicated by a blinking display and/or
faded numbers.

Batteries For Low Temperature Applications

Using standard batteries in low temperature condtions
will shorten the life. Change to 9 volt lithium batteries
(Radio Shack part number 23-665) to get longer life.

REFRIGERANT PLUG-IN

The manifold holds up to 32 refrigerants.
These are contained in a replaceable plug-in
part. This is servicable by manufacture only.

A list of refrigerants included with the stock
manifold is shown on page 4.

Digital Manifold

Superheat/Subcooling, Charging/Testing Manifold
& Operating Instructions

®

HIGH SIDE DISPLAYS

PRESSURE, SUBCOOLING and

TEMPERATURE (T2), and (T1-T2)
DIFFERENCE
PUSH to toggle between displays

PUSH to select refrigerant (32 available)
HOLD DOWN to change units to Metric

or English

PUSH once to turn ON or OFF

PUSH twice to select 1 hr/24 hr mode

Use Temperature Jack (T2) for
Subcooling Testing