Time – Torbal AGC Series Manual User Manual
Page 19

Instruction Manual - TORBAL AGC Series
8. Once the weight is removed the scale will continue with calibration adjustment. CAL will be
displayed during the remainder of the adjustment process. Once finished the scale will automatically
return to the weighing mode.
7.1.1.2.2.Time
Internal calibration can be configured to calibrate automatically within a change of time. To set
a calibration time interval, follow the steps below.
1. Press the “T” (YES) key when “TIME” is displayed in the calibration setup menu.
2. “t-0” and the current time interval setting will begin to scroll on the display sequentially. To
disable time interval calibration press the “T” (YES) key when “t-0” is displayed. To enable time
calibration press the “T” (YES) key when the default time setting is displayed.
To Disable
To Enable
19