background image

Configuring Sending Data Modes – Scale to PC

AD and AGC Models:

1. Press the M key
2. Wait for function “Print” (print) to appear and press the T key.
3. Wait for “Send” (Send) to appear and press the T key.
4. Once a desired sending mode is displayed press the T key to activate.

Stb (Stb)

Stable

Stable weighing result is transferred from the scale after pressing the P key

Auto (auto)

Automatic

Weighing result is automatically transferred from the scale after it
stabilizes (no push key required)

Cont (Cont)

Continuously Weighing results is continuously transferred from the scale at all times. (no

push key and no stability required)


AGCN and AGZN Models:

1. Press the MENU key
2. Wait for function “RS-232” (RS-232) to appear and press the T key.
3. Wait for “Sending” (SendING) to appear and press the T key.
4. Once a desired sending mode is displayed press the T key to activate.

b. P Stb (Stb)

Button plus
Stability

Stable weighing result is transferred from the scale after pressing
the P key,

butt. P (Butt. P)

Button Only

Weighing result is transferred from the scale after pressing the P
key (no stability required)

Auto (auto)

Automatic

Weighing result is automatically transferred from the scale after it
stabilizes (no push key required)

Cont (Cont)

Continuously

Weighing results is continuously transferred from the scale at all
times. (no push key and no stability required)

AG and AGN Models:

1. Press the MENU key
2. Select Configuration
3. Select RS-232C or USB.
4. Use < > keys to select a data sending mode.

button + stab.

Button plus
Stability

Stable weighing result is transferred from the scale after pressing

the

key,

button only

Button Only

Weighing result is transferred from the scale after pressing the

key (no stability required)

after stabil.

Automatic

Weighing result is automatically transferred from the scale after it
stabilizes (no push key required)

continuous

Continuously

Weighing results is continuously transferred from the scale at all
times. (no push key and no stability required)