Vetter Permanent Aspirator

6.

Elimination of defects

Defect

Cause

Remedy

Suction motor

does not run

Shut-off switch is off

Turn the switch on. Call

a qualified electician if

switch cuts off again.

No mains voltage

Check socket

Connection cable defective Replace cable*
Motor is defective

Exchange motor*

Emergency switch is

pressed

Unlock the emergency

switch

Suction motor

runs but equip-

ment does not

suction

Float ball has closed the

suction opening

Switch off motor, re-

lease float ball

Suction line is blocked

Clean suction line

Motor head turbine is

defective

Carry out maintenance*

Conveyor pump

power is too low

Pump is blocked

Clean pump

Conveying height is too

high

Reduce height diffe-

rence

Suction motor

runs but suction

power is too low

Equipment is drawing air

from outside

Check the adapter and

motor head for correct

seating

Sealing is defective

Exchange sealing

Suction line is blocked

Clean suction line

Suction height is too great

Place the equipment to

suction level

Filter basket is blocked

Empty filter basket

Conveyor pump

does not work

Float switch is interlocked

Release float switch

Connection cable defective Replace cable*
Pump wheel blocked by dirt Clean pump and wheel
Pump motor defective

Exchange pump*

Conveyor pump

runs but there is

no feed

Pump blocked

Clean pump

Ventilation hole blocked (in

the lifting tube of the pump

under the flap trap or at the

side of the pump)

Clean ventilation hole

Pressure hose is bent

Take out the bend in

the hose

Shut-off unit Size C is closed Open shut-off unit
Pump is defective

Exchange pump*

*only by a qualified electrician

Should the defect remain despite the above information then contact an authorized

representative or Vetter GmbH.