Appendix, Timer, Message codes – Leica Geosystems Leica DISTO A5 User Manual

Page 11: Technical data

Advertising
background image

8

Appendix

Leica DISTO™ A5 1.3.0 gb

D

GB

F

I

E

P

NL

DK

S

N

FIN

J

CN

RC

ROK

PL

H

RUS

CZ

Timer

Press and hold TIMER - key {A, 3} until desired time
delay has been reached (5 - 60 seconds).
Press DIST - key {A, 1}. Once the key is released the
remaining seconds until measurement are displayed.
The last 5 seconds are counted down with a beep.
After the last beep the measurement is taken.

Special function: Switch off of Beep
Pressing MENU - key {A, 9} and MINUS - key {A, 12}
together for 4 sec. switches the Beep off or on.

Appendix

Message Codes

All message codes will be displayed with either "InFo"
or "Error".
The following mistakes can be corrected:

Technical Data

InFo

Cause

Remedy

204

Calculation error

Repeat procedure

206

No endpiece detec-
tion

Attach the endpiece
properly. If error still
occurs, replace the
endpiece.

252

Temperature too
high

Cool down instrument

253

Temperatur too low Warm up instrument

255

Receiver signal too
weak, measurement
time too long,
distance > 100 m

Use target plate

256

Received signal too
powerful

Use target plate
(grey side)

257

Wrong measure-
ment, ambient
brightness too high

Use target plate
(brown side)

260

Laser beam inter-
rupted

Repeat measurement

Error

Cause

Remedy

Hardware error

Switch on/off the
device several times
and check if the
symbol still appears. If
so please call your
dealer for assistance.

Range (use target plate
for longer distances)

0.05 m to 200 m

0.2 ft to 650 ft

Measuring accuracy up
to 30 m (2

σ, standard

deviation, room tempe-
rature)

typ.: ± 1.5mm*

Smallest unit displayed

1 mm

Laser class

II

Laser type

635 nm, < 1 mW

Ø laser spot (at
distance)

6 / 30 / 60 mm

(10 / 50 / 100 m)

Autom. switch off laser

3 min

Autom. switch off
instrument

6 min