Troubleshooting Guide 29

Troubleshooting Action Table

PROBLEM

ACTION

During battery status
check, no flash or only
one flash is observed.

!" Replace battery
!" If there are no blinks after changing battery,

clean the battery contacts with an alcohol swab.

Red LED indicator
continuously blinks at
one-second intervals –
and/or- Audible Alarm, if
activated, sounds at one-
second intervals.

!" Replace Headpiece cable
!" If problem continues, report immediately to

cochlear implant center. You may need a new
Headpiece or reprogramming.

Continuous rapid blinks
of the red LED indicator.
(May indicate a faulty
program.)

!" Change Program Switch to another program

position.

!" Report problem to parents and/or cochlear

implant center.

Green LED does not
illuminate to loud speech
near microphone.

!" Recheck battery status. If okay, perform next

step.

!" Plug in an auxiliary microphone. If LED lights,

then replace Headpiece.

!" Plug the Headpiece Microphone Tester

Earphones into the external input port of the
Processor and turn the function switch to mic test
(!

).

Note that the LED does not flash in

response to this mode.

!" If problem continues, immediately report to

cochlear implant center.

User reports hearing no
sound.

!" Repeat the system check steps: 1. Battery

Charge Status, 2. Lock Status, 3.
Microphone/System Status.

!" If problem continues, immediately report to

cochlear implant center.

User reports hearing
static

!" Ensure that the cable is inserted into Processor

and that the Headpiece is positioned properly on
the user’s head.

!" Turn the volume and sensitivity controls to 12:00

and 10:30 positions, respectively.

!" If user still reports “static”, replace cable.
!" Visually inspect the microphone for signs of

debris or wear.

!" Use the Headpiece Microphone Tester

Earphones and the microphone tester function.

!" If available, try an auxiliary microphone or

alternate headpiece.

User reports that sounds
are muffled or distorted.

!" Confirm that no clothing or material is

obstructing the microphone opening.

!" Repeat the same steps as when static is reported.

Advertising