Table III-3 Accuracy for LOW TORQUE CAP Viscometer
Cone Number
≤ 500 RPM
10-100% FSR
01
± 2%
02
± 2%
03
± 2%
04
± 2%
05
± 2%
06
± 2%
07
± 2%
08
± 2%
09
± 2%
10
± 2%
Table III-4 Accuracy for HIGH TORQUE CAP Viscometer
Cone Number
≤ 400 RPM
750 RPM
900 RPM
900 RPM
10-100% FSR
10-100% FSR
≤ 50% FSR
> 50% FSR
01
± 2%
± 2%
± 2%
± 4%
02
± 2%
± 2%
± 2%
± 4%
03
± 2%
± 2%
± 2%
± 4%
04
± 2%
± 3%
± 3%
± 6%
05
± 2%
± 4%
± 4%
± 8%
06
± 2%
± 5%
± 5%
± 10%
07
± 2%
N/A
N/A
N/A
08
± 2%
N/A
N/A
N/A
09
± 2%
N/A
N/A
N/A
10
± 2%
± 2%
± 2%
± 2%