Brookfield DV-E Viscometer User Manual

Page 14

Advertising
background image

Brookfield Engineering Labs., Inc.

Page 14

Manual No. M98-350-J0912

cP
12RPM %

S02

(MOTOR ON)

Figure II-5

cP
12OFF %

S02

(MOTOR OFF)

Figure II-6

Note: When the motor switch is in the ON position, any change to the selected speed

will be effective immediately. When collecting data at multiple speeds, you

may wish to leave the SPEED/SPINDLE switch in the left position to facilitate

speed changes. Also, when the motor switch is turned off, the display will hold

the last measured torque value and measured viscosity.

The DV-E Viscometer employs an optical signal pick-up inside the instrument to detect the torque

value of the calibrated spring. This optical signal pick-up is recorded four times per revolution of

the spindle. When the spindle begins to rotate at a defined speed, four torque values are recorded

during the first full revolution of the spindle and averaged together. The display reports the average

value for both torque (%) and viscosity (cP or mPa•s). Thereafter, the next torque value recorded by

the optical signal pick-up is averaged together with the three preceding torque values and the newly

calculated torque (%) and viscosity (cP or mPa•s) values are then displayed. This 4x revolution

pick-up and display continues as long as the motor is on.

This algorithm in the instrument firmware is used for all viscosity and torque readings. Consequently,

the wait time to observe the initial displayed readings for torque and viscosity increase as you go

to lower speeds.

It may also be necessary to allow time for the indicated reading to stabilize.

Note: At speeds of 1 RPM and lower, additional time may be required to allow for

complete deflection of the torque sensor. The % (torque) and cP (viscosity) will

flash until 1 revolution is achieved and the % torque value is greater than 10%.

The time required for stabilization will depend on the speed at which the Viscometer is running

and the characteristics of the sample fluid. For maximum accuracy, readings below 10% should be

avoided. Additional information on making viscosity measurements is available in Appendix B or

the Brookfield publication “

More Solutions to Sticky Problems”

.

The

DV-E

Viscometer will remember the selected speed and spindle when power is turned off. On

start-up, the Viscometer will be set to the previously selected spindle and speed.

Please see Brookfield publication “

More Solutions to Sticky Problems”

(Chapter 3) for information

on how to select a speed.