Brookfield Bath, Model TC-602 User Manual

Page 13

Advertising
background image

12

Sub-Menu Selections — To select an item in a sub-menu, press the Select/Set Knob. The first
available item (or the last item in that sub-menu which was selected) will be highlighted. Rotate the
Select/Set Knob clockwise to advance to the next sub-menu item; rotate the knob counter-clockwise to
go back to the previous item. Press the Select/Set Knob to select the highlighted item; the highlighting
will change from

THIS

to

THIS, indicating that the displayed value or choice may be changed.

Entering and/or Changing Sub-Menu Values — Once the desired sub-menu item has been selected
(as described above), the displayed value is changed by rotating the Select/Set Knob. The change is
accepted by pressing the Select/Set Knob. On sub-menu items requiring multiple entries, such as
hours/minutes/seconds, an underline will appear under the first value in that sequence which can be
changed. (Example: 01:23:00 ) To accept the displayed value and/or advance to the next value in
the sequence, press the Select/Set Knob. To return to the previous cursor position, such as from
minutes back to hours, press the ESC Button. Once a value has been entered and accepted, the
highlighting box around the value will disappear. If you do not wish to accept the displayed value,
press the ESC Button or allow the display to timeout before pressing the Select/Set Knob.

Menu Structure

Main Menu Item

Associated
Sub-Menu Items

Choices / Ranges / Comments

Timer

Set
Beep

00:00:01 to 99:59:59
On or Off

Program

Program #
Program Steps
Program Loops
Step #
Step Setpoint
Minutes/Seconds
View Profile

1 to 10
1 to 50
1 to 99
1 to 50
–50° to 200° (-50° to 392°F)
1 second to 999 minutes, 59 seconds
Displays temperature profile of program.

Run Program

Program #
Program Status

1 to 10
Start, Starting, Running, Paused, or Completed

Preferences

Readout
Units
Sound
Language
Program

1.

#, #.#, #.##, or #.### (0, 1, 2, or 3 decimal places)
°C or °F
1 to 100
English / French / German / Spanish
Time / Temperature

Limits / Alarms

Limit High
Alarm High
Alarm Low
Limit Low

-50° to 200°C (-58° to 392°F)
-50° to 200°C (-58° to 392°F)
-50° to 200°C (-58° to 392°F)
-50° to 200°C (-58° to 392°F)

Pump

Pump speed

Low to High

Temperature
Trend

No sub-menu; displays
temperature profile

2 minutes to 48 hours

Probe

Internal / External
Maximum Setpoint
Differential

Probe currently selected is shown
Only displayed when External is selected
1° to 10°C range

Auto
Refrigeration

1.

Refrigeration On
Setpoint

20° to 85°C

Instrument

Contrast
Timeout
Baud Rate

01 to 20
5 to 60 seconds
110 / 300 / 600 / 1200 / 2400 /4800 / 9600 /
14400 / 19200 / 38400 / 57600

1.

This Main Menu item is present on Refrigerating/Heating Circulators only.

NOTE: There are additional displays after “Instrument “ in the Main Menu. However, there are no
user-settable functions on these displays.