Brookfield Bath, Model TC-602 User Manual
Page 13

12
Sub-Menu Selections — To select an item in a sub-menu, press the Select/Set Knob. The first
available item (or the last item in that sub-menu which was selected) will be highlighted. Rotate the
Select/Set Knob clockwise to advance to the next sub-menu item; rotate the knob counter-clockwise to
go back to the previous item. Press the Select/Set Knob to select the highlighted item; the highlighting
will change from
THIS
to
THIS, indicating that the displayed value or choice may be changed.
Entering and/or Changing Sub-Menu Values — Once the desired sub-menu item has been selected
(as described above), the displayed value is changed by rotating the Select/Set Knob. The change is
accepted by pressing the Select/Set Knob. On sub-menu items requiring multiple entries, such as
hours/minutes/seconds, an underline will appear under the first value in that sequence which can be
changed. (Example: 01:23:00 ) To accept the displayed value and/or advance to the next value in
the sequence, press the Select/Set Knob. To return to the previous cursor position, such as from
minutes back to hours, press the ESC Button. Once a value has been entered and accepted, the
highlighting box around the value will disappear. If you do not wish to accept the displayed value,
press the ESC Button or allow the display to timeout before pressing the Select/Set Knob.
Menu Structure
Main Menu Item
Associated
Sub-Menu Items
Choices / Ranges / Comments
Timer
Set
Beep
00:00:01 to 99:59:59
On or Off
Program
Program #
Program Steps
Program Loops
Step #
Step Setpoint
Minutes/Seconds
View Profile
1 to 10
1 to 50
1 to 99
1 to 50
–50° to 200° (-50° to 392°F)
1 second to 999 minutes, 59 seconds
Displays temperature profile of program.
Run Program
Program #
Program Status
1 to 10
Start, Starting, Running, Paused, or Completed
Preferences
Readout
Units
Sound
Language
Program
1.
#, #.#, #.##, or #.### (0, 1, 2, or 3 decimal places)
°C or °F
1 to 100
English / French / German / Spanish
Time / Temperature
Limits / Alarms
Limit High
Alarm High
Alarm Low
Limit Low
-50° to 200°C (-58° to 392°F)
-50° to 200°C (-58° to 392°F)
-50° to 200°C (-58° to 392°F)
-50° to 200°C (-58° to 392°F)
Pump
Pump speed
Low to High
Temperature
Trend
No sub-menu; displays
temperature profile
2 minutes to 48 hours
Probe
Internal / External
Maximum Setpoint
Differential
Probe currently selected is shown
Only displayed when External is selected
1° to 10°C range
Auto
Refrigeration
1.
Refrigeration On
Setpoint
20° to 85°C
Instrument
Contrast
Timeout
Baud Rate
01 to 20
5 to 60 seconds
110 / 300 / 600 / 1200 / 2400 /4800 / 9600 /
14400 / 19200 / 38400 / 57600
1.
This Main Menu item is present on Refrigerating/Heating Circulators only.
NOTE: There are additional displays after “Instrument “ in the Main Menu. However, there are no
user-settable functions on these displays.