2 other hazards, 3 safety measures, 4 safety elements – BUCHI Rotavapo R-215 User Manual
Page 9

2 Safety
9
R-210/215 Operation Manual, Version F
WARNING
Potential explosion risk if solvent vapors accumulate within the instrument housing.
• Always use the instrument in a well ventilated area.
• Beware of damaged or cracked glass parts.
• Beware of the fire hazard.
2 .5 .2
Other hazards
WARNING
Certain solvents in or in the vicinity of the Rotavapor can form peroxides and/or are highly inflam-
mable.
• Always be aware of the explosion risk when working with hazardous substances or with
substances of unknown composition.
• Always provide a good ventilation within or in the vicinity of the system.
2 .5 .3
Safety measures
Always wear personal protective equipment such as protective eye goggles, protective clothing and
gloves when working with the instrument.
2 .5 .4
Safety elements
Electronics
• The heating bath is equipped with a mechanical and an electronic over-temperature protection.
The mechanical over-temperature protection consists of a bimetal thermostat that, in case of over-
temperature (over 260 °C), directly interrupts the power supply. It has to be set back manually after
the bath has cooled down (see also chapter 8).
The electronic over-temperature protection controls the temperature limit (actual bath temperature
may not exceed the given temperature by 2 °C for more than 2 minutes), the heating rate (actual
temperature may not rise by more than 5 °C during 5 seconds) and the function of the temperature
sensor.
• The heating bath is equipped with safety fuses.
• Electronic control of the heating bath temperature - to prevent the product from overheating
Parts in direct contact with the instrument
• Safety catch for adjusting the immersion depth of the evaporating flask into the heating bath.
• Combi Clips for fixing the evaporating flask and for safe removal of fixed ground-glass joints.
• Ball joint clip for safe fixing of the receiving flask.
• Rods and holders for fixing the glass assemblies.
• Electronic over-current protection at the drive unit and lift motor.
• Safety spring preventing the vapor duct from dropping out.
• Automatic raising of the flask from the heating bath in case of a power failure.
Glass
• Use of high quality, inert 3.3 borosilicate glass.
• Use of tube clips GL-14 for preventing glass breakage.