Danger – BUCHI Nano Spray Dryer B-90 User Manual

Page 11

Advertising
background image

2 Safety

11

B-90 Operation Manual, Version B

!

Danger

Death by suffocation or serious poisoning by inhalation of inert gases.
• Do not inhale inert gases
• Directly withdraw released gases and gaseous substances by sufficient ventilation
• Only operate the instrument in ventilated environments
• Before operation, check all parts, connections and sealings involved in the gasflow for proper

sealing

• Exchange worn out or defective parts immediately

!

Danger

Death or serious poisoning by gases or particles due to O

2

– sensor or filter malfunction

• Exchange defective O

2

– sensor immediately

• Exchange O

2

– sensor regularly within the specified maintenance intervals

• Exchange clogged filters immediately
• Exchange filters regularly within the specified maintenance intervals
• Dispose of filter immediately

!

Danger

Death or serious poisoning by inhalation or incorporation of dried particles during spray process.
• Wear safety goggles
• Wear safety gloves
• Wear a suitable protective mask
• Wear a laboratory coat
• Check for proper sealing before use
• Do not inhale dried particles
• Stop drying gas flow before opening the drying circuit

Advertising