BUCHI Melting Point M-560 User Manual

Page 33

Advertising
background image

6 Operation

33

Melting Point M-560 Operation Manual, Version D

• The instrument auto-
matically heats to the start
temperature of the first
substance. In the meantime,
prepare at least 6 samples of
the given substance. For this
process, follow section 6.2.1 .

• Choose the correct phar-

macopoeia melting point
value with the rotary knob.
The pharmacopoeia melting
point is written on the
certificate of analysis. Press
Save.

• To register temperature

stamps manually press the
Set button for the related
sample. Each button can
be pressed once. Regis-
tered temperature stamps
are displayed green.

After each run the current
result is indicated.
• Press

START to perform

next measurement.

After 6 results are obtained
within a standard deviation of
+/- 0.2 °C, the process moves
on to the next substance. The
process is identical for the
subsequent substances.

As soon as a complete
calibration is obtained, the
new calibration date is saved
automatically.
• Press

OK. All results

obtained are displayed.

• The rotary knob can be

used to display all the
results of the calibration.

• The

Print key is used to

print out the data.

Back exits this menu.

The calibration date is now
saved and indicated on the
idle screen.