5 Putting into operation

18

Pump Combinations Operation Manual, Version E

5 .3 Installation

In order to make things clearer, the following descriptions do not contain the complete names of the
modular pump systems, but rather the general name.

Pump Controller C-610: Pump controller

Pump Manager C-615: Pump manager

Pump Module C-601/605: Pump module

Mixer (pressure sensor with integrated mixing chamber): Pressure sensor

Control Unit C-620: Control unit

5 .3 .1

Installating the standard accessories

Suction tubing 4 mm x 2.5 mm, length 1.5 m

ATTENTION
The pump module may never be operated without the suction filter, since otherwise impurities may
enter the pump system.

1

2

3

Cut the suction tubing a to the desired length and

fasten it to the suction filter. Place the fitting b and

the ferrule c on the other end of the suction tubing
a. Attach the fitting b and the ferrule c.

Press the suction tubing

a and the ferrule c against

a table surface to firmly seal the tubing end with
the ferrule.

Fig. 5.1: Installing fitting and ferrule

2

1

4

Remove the protective foil from the input and output
of the pump head. Screw the white fitting b into the

lower input d of the separator head. Make sure that

all fittings are hand-tight. The end of the tubing a

with the solvent filter goes into the solvent vessel.

Fig. 5.2: Pump head with connected suction tubing