8 troubleshooting, 1 malfunctions and their remedy – BUCHI Wet Digester B-440 User Manual

Page 31

Advertising
background image

8 Troubleshooting

31

B-440 Operation Manual, Version D

8 Troubleshooting

This section helps to resume operation after a problem has occurred with the instrument which does

not require special technical training. It lists possible occurrences, their probable causes and suggests

how to remedy the problem.

The troubleshooting table below lists possible malfunctions and errors of the instrument. The operator

is enabled to correct some of those problems or errors by him/herself. For this, appropriate corrective

measures are listed in the column “Remedy”.

8.1

Malfunctions and their remedy

Malfunctions and their remedy

Malfunction

Possible cause

Remedy

Heating performance low

Blind reflector or worn out

heating lamp (see section 8.1.1)

Exchange reflector or call service for

bulb exchange

LED display does not illuminate

Onboard fuse or electronic

malfunction

Call service for repair

Heating lamp does not warm up

Bulb faulty

Call service for exchange

E2, E4, E5, E6,E7, E8, E9

Electronic malfunction

Call service for repair

Malfunctions or errors which are not listed in the table are to be corrected by a Buchi trained technical

engineer who has access to the official service manuals. In this case, please refer to your local Buchi

customer service agent.

8.1.1

Heating performance test

Exchange the reflector if a temperature of 530 °C can no longer be achieved at the bottom inside the

crucibles. To check this, a Tempilstick® with a melting point of 530 °C can be used.

NOTE

If the reflector exchange does not increase the system performance sufficiently, call service for a

heating bulb exchange!

Advertising