5 Putting into operation

23

NIRMaster Operation Manual, Version A

NOTE

To cut the power in case of an emergency by unplugging, the instrument or any other item must

•

not block the mains plug! In this case, the plug must be able to be pulled out instantly.

When the system is installed with a locked or secured plug an external power switch is required.

•

When the system is installed stationary (without power plug) an external power switch is

•

required.

External power switches (e.g. emergency stop switches) must meet IEC 60947-1 and IEC

•

60947-3 requirements. Such devices must be clearly labeled and accessible at any time.

Additional electrical safety measures such as residual current breakers may be necessary to meet

•

local laws and regulations!

External connections and extension lines must be provided with a grounded conductor lead

•

(3-pole couplings, cord or plug equipment). All used power cords must meet the input power
requirements.