NOTE:

The Instruction panel will instruct you to insert an empty vial in
order to perform the prewash

4.

Raise the empty vial to enable the Start button

5. Click

the

Start button

NOTE: After the prewash is complete, the
instrument will then obtain the voltages
for each sensor at each of the specified
temperatures.

NOTE: The instruction panel will then
instruct you to insert a vial containing an
oil with a viscosity 3 times the lower limit
of the tube and raise it into the ready
position for determining trip points.

NOTE:

Once all voltages and trip points

have been obtained for all desired
temperatures, the tube sensors have been
trained.










Bath fluid safety features:

If the bath fluid level drops to an
unacceptable level, the light in the bath unit
will go out, alerting the user that the level
may be too low. Bath operations should be
aborted until the level is checked and if
needed, restored to an acceptable bath
fluid level.

Checking the bath fluid level:


1.

Open the left side panel (left ear) to
view the LED light and to check the
bath fluid level.