Viewing/editing setup information

NOTE

For first-time installation, make certain that the factory-prepared Configura-
tion file has been copied to the Default Data directory per instructions on
page 7 of manual. If your instrument has already been configured, you can
use the instructions in this section of the manual to check or, if necessary,
change the instrument settings.

1. Click

Configure

from the VISCPRO

®

menu bar.

2. Select your instrument group and instrument from the list of available

instruments.

3. Select

Tray Settings: Tube and Bath

from the list of configuration

options.

The Tray Settings: Tube and Bath

window will appear.

The Tray Settings: Tube and Bath window

The

Tray Settings: Tube and Bath

window contains setup information

for each tube associated with your instrument. On multiple-tube CAN-
NON instruments, you can click on the tube tabs to see the setup infor-
mation for different tubes.

4. Verify that the tube serial number (

Tube S/N

) is correct. If it is not,

input the correct serial number in the text box. Each tube should have
a unique serial number!

5. Verify the presence of calibration

values (

C

and

E

) for each bulb. If

calibration data is not available, the
default values are C=1 and E=0.

Sample calibration constants