Checkline MCT-HS User Manual

Page 5

Advertising
background image

Electromatic Equipment Co., Inc.

-12-

17.0 MOST COMMON REASONS FOR MISS READINGS

•

Product temperature out of application range
Material below 0°C resp. above +40°C may cause faul ty measurements. The
storage of cold material in a warm storage area usually creates condensed
water which may lead to major measuring errors.

•

Discrepancy in temperature between device and material
Please ensure that the device and the material under test are being stored at the
same temperature (+/- 3°C) before measuring. Protec t your measuring device
from direct sunlight for a reasonable time period before taking a measurement.
A high temperature difference has a negative effect on the stability of the
measurement results.

•

Wrong calibration curve
Double check the correct selection of the calibration curve before measuring.

•

Wet or mouldy material

•

Frozen measuring material or material containing snow
This leads to a major decrease in accuracy.

•

Direction of plugging in
The Direction of plugging has a great influence on the accuracy. Necessarily
follow the instructions according to the beginning of the manual!

•

Compressed density outside the application range
If the compressed density differs from that specified, there may be deviations!

•

Moving the measuring head after the plug in leads to miss readings!

•

Water film at the measuring head
After measuring wet material a water film can arise on the sensor head. This
could lead to a too high result in the following measurements. After measuring
wet material clean both black plastic parts of the measuring head accurately with
a dry cloth.



ATTENTION: Risk of injury by measuring head!
Keep away from children younger than 16 years!

Electromatic Equipment Co., Inc.

-5-

5.0

MENU LEVEL OVERVIEW





































Edit Logs

Manual Logs

Clear Logs


Print Logs

Last Log

All Logs

Clear Logs


Send Logs

Manual Logs

Clear Logs

Options

Date / Time

Log Time

Language

Unlock

°C / °F

o Userlevel

BL On

Time

Auto Off Time

o Online Send

o Online Print

Materialcalib.

Password

Reset

Type selection

Save new value (log)
Power off (3 sec.)

Watch the saved data (logs)

Main menu

Switch lower
Open this menu / Enter

Switch upper

Overview main menu

Previous calibration curve
Power off (3 sec.)

Next calibration curve

Advertising