OTHER INSTRUMENT FUNCTIONS – OVERVIEW


•

Manual saving of single measurement values in a measurement series

•

Automatic saving of measurement values in a measuring series

•

Selection of the automatic saving interval (3 seconds to 4 hours)

•

Display of the measuring series and measurement values directly on the

instrument

•

Printout of the saved measuring series

•

Transfer and saving of the measuring series on a PC

•

Display of the memory and battery status

•

Selection of the menu language (DE, EN, FR, IT)

•

Temperature shown in degrees Celsius or degrees Fahrenheit

•

Acti vation of a simple user operation

Electromatic Equipment Co., Inc.

-9-

7.0

COMMON REASONS FOR INCORRECT MEASUREMENTS



Sunlight or other sources of heat or cold which do not correspond to the
ambient temperature

Measuring errors due to differences in temperature

To show how important it is that the temperature of the instrument is the same as
the temperature of the substance to be measured, the table below shows measuring
errors due to a temperature difference of only 1°C / 1.8°F between the measuring
instrument and the substance to be measured at different ambient temperatures.

10°C (50°F) 20°C (68°F) 30°C (86°F)

10%r.H. ± 0.7%

± 0.6%

± 0.6%

50%r.H. ± 3.5%

± 3.2%

± 3.0%

90%r.H. ± 6.3%

± 5.7%

± 5.4%


At room temperature (20°C/68°F) and assumed paper m oisture value of 50%r.H. a
deviation of 1°C / 1.8°F between the measuring sens or and the substance to be
measured results in a measuring error of 3.2%r.H. A deviation of 3°C / 5.4°F would
result in a measuring error of over 10%.

Further examples are given in the „Mollier i-x“ diagram.

Dripping or sprayed water

Irreversible damage to the sensor element due to aggressive gases

Danger of condensation following temperature changes

Dirty moisture sensor

Foreign objects on the sensor