4.0

CONTACT-FREE TEMPERATURE SENSOR


The back side of the humimeter PM4 contains
an embedded temperature sensor for the
contact-free temperature measurement below
the sensor. This ensures the determination of
the temperature of the product within seconds
and

consequently

the

temperature

compensation of the displayed water content.
The sensor must not be covered by a hand or
any other object!


5.0

CALIBRATION CURVES


name

paper type

density

600 paper

Very low density paper

600 kg/m³

650 paper

Low density paper

650 kg/m³

700 paper

Low density paper

700 kg/m³

750 paper

Fluting, Schrenz

750 kg/m³

800 paper

Newsprint paper

800 kg/m³

850 paper

Kraftliner brown

850 kg/m³

900 paper

Kraftliner White Top, Testliner brown

900 kg/m³

950 paper

Testliner white, copy paper

950 kg/m³

1000 paper

Copy paper, LWC raw

1000 kg/m³

1050 paper

Copy paper calendered

1050 kg/m³

1100 paper

Copy paper calendered

1100 kg/m³

1200 paper

LWC calendered

1200 kg/m³

1300 paper

Flysheet paper

1300 kg/m³

1400 paper

Flysheet paper

1400 kg/m³

1500 paper

Flysheet paper high density

1500 kg/m³

1600 paper

Flysheet paper very high density

1600 kg/m³

Empty 1700

For creation of customer calibration curve

Empty 975

For creation of customer calibration curve

Empty 550

For creation of customer calibration curve

reference

! Only for checking the instrument !


6.0

LIST OF CALIBRATION CURVES



By pressing one of the arrow keys in the measuring
window for approx. 3 seconds a list of all available
calibration curves opens. Now you can select your
desired curve using the arrow keys and confirm by
pressing .

Electromatic Equipment Co., Inc

-13-

17.0 COMMON REASONS FOR INCORRECT MEASUREMENTS

Wrong calibration curve
Double-check the correct selection of the calibration curve before measuring.
The calibration curve „reference“ must NOT be used.

Inadequate thickness of the paper pile

A single sheet of paper must NOT be measured. Make sure that the paper pile
below the steel clamps has a thickness of at least 100 mm.

Electro conductive material

Every metallic object as well as electro conductive packing material e.g. paper
coloured by soot or wire has a negative effect on the accuracy of your
measurement. Make sure that in an area of 20 cm below the sensor steel
clamps there is no such material.

Product temperature out of application range

The application range is between 0 and +60°C.

Discrepancy in temperature between device and material

Please ensure that the device and the material under test are being stored at
nearly the same temperature before measuring. A high temperature difference
has a negative effect on the stability of the measurement results.

Measurement at the front side

A measurement at the front side of a paper roll will bring different measuring
results. For that you have to select a different calibration curve.

Bended steel clamps

If the steel clamps (measuring area) are bended you might get wrong measuring
results.


Please note:

After measuring at the running paper roll for a longer period the steel clamps will
get hot – don´t touch! => risk of getting burned!

Measuring fine paper at the running roll may cause pressure marks on the paper
=> for possible damages at the paper we refuse any liability.