7 Maintenance of NucleoCounter®

991-0251 Rev. 1.0

Feb-2013

18

The illustration above shows how to wipe the optical window. Place a clean Alpha Swab at

the inner part of the window, and then move it outwards and repeat t o cover the complete

optical window.

An object on the surface of an optical component can influence the collected image. A

contaminant will normally be visible as a faint object in the image. In case the

contamination is on the Slide it will not be visible when a new Slide is analyzed. If the

contamination is on the optical window it will be present on every image taken at this

view

4

of the optical window. If the contamination is present on every image taken no

matter which view is used it is most likely that other optical parts are contaminated. In

this case contact ChemoMetec A/S. An example of a relatively large contamination is given

in Figure 12.

Figure 12. An image showing a contamination of the optical system, visible as a white

cloudy phenomena in the upper right hand corner.

The presence of a contaminant (e.g. dust particles) will normally not influence the

counting of cells. The NucleoCounter® NC-250™ will distinguish between cells and

contaminants, since a cell generally is significantly smaller than contaminants.

7.1.4

Spill of liquid

If liquid has been spilled on the instrument it can contaminate elements of the optical

system. This contamination can show up in the image in several ways and have many

causes but a common feature is that it cannot be removed by the use of compressed air.

Even though such phenomena are clearly visible on the image it only rarely affects the

results of the NucleoCounter® NC-250™. Since it is possible to damage the optical system

while attempting to clean the system it is recommended that ChemoMetec A/S should be

consulted before the user attempts such cleaning.

7.2

Decontamination

In case decontamination of the NucleoCounter® NC-250™ is required please contact

ChemoMetec A/S before taking this action.

4

A slide can have as many as 40 views.