21 en, Emc table – Compex SP4.0 User Manual

Page 23

Advertising
background image

21

EN

8. EMC TABLE

The Compex stimulator needs special precautions regarding EMC and needs to be installed and put into
service according to the EMC information provided.

Portable and mobile RF communications equipment can affect the Compex stimulator.

The use of Accessories, transducers, and cables other than those specified by the manufacturer, may result
in increased Emissions or decreased Immunity of the Compex stimulator.

The Compex stimulator should not be used adjacent to or stacked with other equipment and if adjacent
or stacked use is necessary, the Compex stimulator should be observed to verify normal operation in the
configuration in which it will be used.

GUIDANCE AND MANUFACTURER’S DECLARATION – ELECTROMAGNETIC EMISSIONS

The Compex stimulator is intended for use in the electromagnetic environment specified below. The customer or the user of

the Compex stimulator should assure that it is used in such an environment.

EMISSION TESTS

COMPLIANCE

ELECTROMAGNETIC ENVIRONMENT

– GUIDANCE

Radiated Emissions CISPR 11

Group 1

The Compex uses RF energy only for
its internal function. Therefore, its RF

emissions are very low and are not

likely to cause any interference in

nearby electronic equipment.

Conducted Emissions CISPR 11

Class B

Harmonic emissions IEC 61000-3-2

Class A

The Compex is suitable for use in all

establishments, including domestic

establishments and those directly

connected to the public low-voltage

power supply network that supplies

buildings used for domestic purposes.

Voltage fluctuations IEC 61000-3-3

Complies

Advertising