2

•

•

•

•

•

•

Front Panel (optional)

•

Monochrome 480 x 272 LCD display with backlight

•

Five tactile keys for menu item selection

Load Cell Excitation

•

5 VDC +/- 1.15 VDC maximum

•

Isolation from digital section 1000 VDC minimum

C2 Calibration Input

•

Isolation from digital section 1000 VDC minimum

Cable Lengths

•

250 feet maximum of C2 authorized cable (Maximum of 4 load sensors) with IT

Junction box

Network Connectivity

Ethernet TCP/IP

Ethernet UDP

Modbus-RTU over RS485

HI 6310 Only
Analog 4-20 mA

Printer port

USB port

Environmental Requirements

Operating Temperature Range

•

0º to 60º C (32º to 140º F)

Temperature Coefficient

•

Less than 0.005% of full scale per degree C for Cal-LO and Cal-HI reference points

Storage Temperature Range

-40 to 85º C (-40º to 185º F)

Environmental

•

Intended for Building In, indoor use only at ambient temperatures between 10ºC to

60º C (14º to 140º F) with a pollution degree of 2.

Humidity Range

•

0-90% (non-condensing)

Maximum Altitude for Installation

•

2000 Meters (6,562 Feet)