Care and maintenance, Limited warranty – American Diagnostic Corporation (ADC) Diagnostix 952B User Manual

Page 4

Advertising
background image

4. Position the Stethoscope

Position the chestpiece in the antecubital space below the cuff, distal to the brachium. Do not place chestpiece underneath

the cuff, as this impedes accurate measurement. Use the bell side of a combination stethoscope for clearest detection of the

low pitched Korotkoff (pulse) sounds.

5. Deflate the cuff

Open the valve to deflate the cuff gradually at the AHA recommended rate of 2-3mmHg per second.

6. Measurement

Record the onset of Korotkoff sounds (Phase 1) as the systolic pressure, and the disappearance of these sounds (Phase 5) diastolic

pressure. (Some healthcare professionals prefer recording diastolic 1 and diastolic 2. Diastolic one occurs at phase 4).

After measurement is completed, open valve fully to release any remaining air in the cuff. Remove cuff.

Care and Maintenance

Storage

After measurement, wrap cuff around bulb and store in cuff basket.

Manometer

Periodically replace the filter plug and diaphragm, and clean the inside of the plastic cartridge tube using alcohol and a lint

free pipe brush. Units should be cleaned and serviced periodically depending on frequency of use, or whenever inside of cartridge

tube appears dirty. Proper servicing of your mercury unit according to a strict maintenance schedule will assure maximum measurement

accuracy and extend the life of your instrument.

For directions on disassembling the unit refer to the enclosed schematic diagram and accompanying instructions.

PLEASE NOTE: MERCURY IS TOXIC. SECURE MERCURY WITHIN THE RESERVOIR BEFORE SERVICING UNIT. ALWAYS

WEAR PROTECTIVE CLOTHING (GLOVES AND GOGGLES) WHEN CLEANING OR SERVICING THE UNIT. CLEAN ANY

SPILLED MERCURY PROMPTLY USING APPROVED MERCURY CONTAINMENT KITS. ONLY PROPERLY TRAINED SERVICE

TECHNICIANS SHOULD ATTEMPT TO SERVICE THE UNIT. CONSULT THE ENCLOSED MANUAL ENTITLED “JUST IN CASE”

FOR PROCEDURES ON DEALING WITH A MERCURY SPILL.

Limited Warranty

American Diagnostic Corporation (ADC) warrants its products against defects in materials and workmanship under normal use

and service as follows:

1. Warranty service extends to the original retail purchaser only and commences with the date of delivery.

2. The entire sphygmomanometer is warranted for one year.

3. The inflation system is warranted for three years.

4. The calibrated plastic cartridge tube is warranted for life and will be replaced FREE OF CHARGE when returned freight

prepaid.

What Is Covered: Calibration, repair or replacement of parts, and labor.

What is Not covered: Transportation charges to and from ADC. Damages caused by abuse, misuse, accident, or negligence.

Incidental, special, or consequential damages. Some states do not allow the exclusion or limitation of incidental, special or con-

sequential damages, so this limitation may not apply to you.

To Obtain Warranty Service: Send item(s) postage paid to ADC, Attn: Repair Dept., 55 Commerce Drive, Hauppauge, NY

11788. Please include your name and address, phone no., proof of purchase, a brief note explaining the problem, and $2.00 to

cover the cost of return shipping and handling.

Symbol

Definition

Important Warning/Caution


Latex-Free


Circumference Size

Conforms to EU Standards

Symbol

Definition

Authorized European

Represenative’s Information

Manufacturer’s Information

Temperature Limit

Humidity Limitation

°