American Diagnostic Corporation (ADC) 421 Adtemp User Manual

Page 6

Advertising
background image

Blank display: Please check if the battery has been loaded correctly. Also check po-

larity (+ and -) of batteries.

Dead battery indication: If the steady battery icon is the only symbol shown on the

display, the batteries should be replaced immediately.

9. Cleaning and Disinfecting

Use an alcohol swab or cotton tissue moistened with alcohol (70% Isopropyl) to clean the

thermometer casing and the measuring probe. Ensure that no liquid enters the interior of

the thermometer. Never use abrasive cleaning agents, thinners, or benzene for cleaning and

never immerse the instrument in water or other cleaning liquids. Take care not to scratch

the surface of the probe lens and the display.

10. Battery Replacement

This instrument is supplied with one lithium battery, type CR2032. Replace with a new

CR2032 battery when the flashing battery symbol appears on the display. Remove the bat-

tery cover by sliding it downward. Remove the battery and replace with a new one.

11. Warranty

This instrument is covered by a 2 year warranty from the date of purchase.

• The warranty covers the instrument and the batteries. The packaging is not in-

cluded.

• Opening or altering the instrument invalidates the warranty.

• The warranty does not cover damage caused by improper handling, discharged bat-

teries, accidents or non-compliance with the operating instructions. Contact ADC® Cus-

tomer Service Dept. for more information.

12. Technical Specifications

Type:

Ear Thermometer 421

Measurement

Range:

32.0°F to 212.0°F - 0°C to 100.0°C

Resolution:

0.1°F/°C

Measurement:

Laboratory: ±0.4°F, 86.6 ~ 108.0°F

Accuracy:

±0.2°C, 32.0 ~ 42.2°C

Display:

Liquid Crystal Display, 4 digits plus special icons

Acoustic:

The unit is turned ON and ready for the measurement:

1 short beep

Complete the measurement: 1 long beep

System error or malfunction: 3 short beeps

Fever alarm: 10 short beeps

Memory:

Auto-Display the last measured temperature

12 readings recall in the Memory Mode

Operating temperature:

41.0°F to 104°F (5°C to 40°C)

Advertising