Specifications and performance – Boekel Scientific Analog Dry Bath Incubators User Manual

Page 2

Advertising
background image

Warranty: when used in laboratory conditions and according to these operating instructions, Boekel warrants this product to be
free of defective material and workmanship for a period of two years from the date of shipment. The liability of Boekel for any
defective equipment during the warranty period shall be limited to the repair of such equipment or replacement thereof without
charge for parts or labor.

SERVICE: RETURN AUTHORIZATION IS REQUIRED.
Should service be required, contact your salesperson or call Boekel Scientific Customer Service at:
215-396-8200. Units returned for repair must be in clean condition.

SPECIFICATIONS AND PERFORMANCE

Dimensions:

Optional Heating

Accommodates

1 Block Model 112001

8” w x 9.8” d x 3.5” h

Blocks

112001-2

(20.3 x 24.9 x 8.9 cm)

110006

30 tubes up to 6mm

2 Block Model 112002

8” w x 9.8” d x 3.5” h

110010

20 tubes up to 10mm

112002-2

(20.3 x 24.9 x 8.9 cm)

110013

20 tubes up to 13mm

4 Block Model 112004

8” w x 13” d x 3.5” h

110016

6 tubes up to 25mm

112004-2

(20.3 x 33 x 8.9 cm)

110020

6 tubes up to 25mm

Standard Block

3” w x 3 ¾ ” d x 2” h

110025

30 tubes up to 6mm

(7.6 x 9.5 x 5.1 cm)

110035

Combination:

Shipping Weight:

3 tubes up to 25mm

1 Block Model 112001/112001-2

6.0 lbs. (2.7 kg)

5 tubes up to 13mm

2 Block Model 112002/112002-2

6.0 lbs. (2.7 kg)

6 tubes up to 6mm

4 Block Model 112004/112004-2

7.0 lbs. (3.2 kg)

110040

1.5 microcentrifuge

Standard Block

2.0 lbs. (1.0 kg)

tubes by 24 places

Power:

110045

0.5 ml tubes by 24 places

1 Block Model 112001

115 VAC, 100 watts

110048

0.2 ml tubes by 24 places

1 Block Model 112001-2

230 VAC, 100 watts

110051

96 well Microplate

2 Block Model 112002

115 VAC, 150 watts

110096

0.2 ml PCR tubes

2 Block Model 112002-2

230 VAC, 150 watts

110099

1.5 ml “dolphin” tubes 24 plcs.

4 Block Model 112004

115 VAC, 300 watts

110002LID*

Lid for 112001 and

4 Block Model 112004-2

230 VAC, 300 watts

112002 Dry Baths

Temp Range:

110004LID*

Lid for 112004 Dry Baths

1 Block Model 112001/112001-2

ambient +5

o

C to 150

o

C

2 Block Model 112002/112002-2

ambient +5

o

C to 130

o

C

* Temp Limited

4 Block Model 112004/112004-2

ambient +5

o

C to 130

o

C

Uniformity:

±0.50

o

C @ 37

o

C

Stability:

1 Block Model 112001/112001-2

±0.45

o

C @ 37

o

C

2 Block Model 112002/112002-2

±0.45

o

C @ 37

o

C

4 Block Model 112004/112004-2

±0.50

o

C @ 37

o

C

Temperature Stability was determined by using NIST
traceable thermometers placed in 16MM glass test tubes,
with oil, placed in standard 16MM aluminum blocks. The
thermometers were suspended in the test tubes. The
stability numbers shown above are an average of two
readings for the Dry Bath Incubator.

Definitions:

Uniformity
is the variation in temperature in the block at
any instant in time.


Stability is the variation in temperature at a single point
over a long period of time.

Note: The temperature variation that is obtained when
using thermometers placed in the thermometer hole will be
greater than listed above. The oil retards the overall
temperature swing of the block. This method of testing was
used since it is the method that has been used historically
on Boekel Analog Dry Bath Incubators.